FAIL hona system-working ka proof hai — assessor ye dekhna chahta hai ki FAIL par aap kya karte ho.
RECALL SOP (IPC)
har pouch par autoclave batch no + date hai → last successful test ke baad ke saare loads identify karo → chair-side cabinets se withdraw karke freeze karo → machine check/repair → successful TST + BI ke baad hi re-sterilise karke wapas issue karo. Agar koi instrument patient par use ho chuka → patient ko notify + post-procedure surveillance (outbreak SOP).
LOG HONESTLY
Registers → Autoclave register → us cycle ki entry Smart Complete se karo — quick chips me 'Test strip: FAIL' bana hua hai, saath me signature pad (autoclave signature-required kind hai). Fir corrective cycle ki PASS entry alag se.
CAPA
FAIL ka root cause (loading pattern, gasket, temperature) + engineer visit + re-validation — CAPA register me entry with evidence.
ASSESSOR KO
recall SOP printout + freeze kiye pouches + FAIL→PASS dono entries dikhao — ye textbook closure hai.
PREVENTION
daily TST habit (tst-test register schedule) — FAIL jaldi pakda jaata hai, kam loads freeze karne padte hain.