Diagnosis: Spore test (weekly biological indicator) fail matlab autoclave ne spores maare nahi — us cycle ka load sterile nahi hai.
Turant karo:
- Us machine ko band/quarantine karo, 'Do Not Use' tag lagao.
- Fail hone wale cycle ke saare instruments (agar use nahi hue) recall karke dobara sterilize karo — ho sake to doosri machine se.
- Naya BI + control strip se re-test karo. Do consecutive PASS aane tak machine use mat karo.
App me: NABH Hub → Daily Logs → Registers → 'spore-test' kind me entry ko FAIL mark karo (result field). Yahi register audit me proof hai ki aapne pakda aur act kiya. Phir NABH Hub → Registers → 'capa' (Corrective + Preventive Action) me ek CAPA khol do — root cause (loading galat / chamber fault / calibration due), immediate action (recall + re-sterilize), corrective (service engineer bulao) aur preventive (weekly BI + har cycle TST strip). AI se draft ke liye Incidents → 'Suggest CAPA' bhi use kar sakte ho.
Aage se rokna: Har cycle par TST/Bowie-Dick strip (tst-test register) + har week BI (spore-test). RegisterSchedule weekly reminder set kar do taaki miss na ho.
Edge case: Assessor FAIL entry dekh ke khush hota hai (matlab system kaam kar raha) — chhupao mat. Chhupana ya blank chhodna zyada bada NC hai.