Assessor sterilization aur autoclave ka kya poochta hai — SOP-STERIL kya kehta hai?

IPC.6 pe CSSD-style flow chahiye: receive→wash→pack→sterilise→store→issue, autoclave 121°C/15 psi, zone-marked room, Spaulding classification.

Where in ORALZY: Sidebar → Setup → NABH Hub → SOP Library → IPC.6 (SOP-STERIL) + ORALZY → Compliance → Sterilisation Log

IPC.6 (Sterilisation) me assessor room ki zoning aur SOP-STERIL flow dono dekhta hai. Room me floor labels: RED zone (receive + wash) → YELLOW (pack + sterilise) → GREEN (sterile storage, closed cabinet, FEFO, max 21 din pouches me). One-door entry, opposite exit — unidirectional flow, clean aur dirty kabhi mix nahi. SOP-STERIL steps: enzyme pre-soak 5 min → ultrasonic clean 10 min → rinse → dry → inspect → pack with Class IV chemical indicator + autoclave indicator tape → autoclave 121°C/15 psi/30 min wet-pack cycle → cool → sterile storage (pouch pe sterilised date + expiry). Spaulding classification zaroor bolna aata ho: Critical (tissue-penetrating instruments) → sterilise; Semi-critical (mouth mirror, impression tray) → high-level disinfect ya sterilise; Non-critical (BP cuff) → low-level disinfect. Flash cycle sirf exceptional case me, director approval + log ke saath. READY: zone labels lage ho, PPE rack entry pe, sterilisation in-charge (ICN-trained) named ho.

sterilizationautoclavecssdspauldingzoningsafaisop-sterildisinfectionaudittaiyariipc-6reprocessing